Understanding Star Switching

Switching probably is the most overlooked and undervalued part of a test system design. Great attention is spent selecting the measurement and stimulus instruments. But more often than not, the signal switching solution does not complement the instruments. It doesn’t matter how accurate the instruments are if the signals pass through a poor switch to get to them.

Engineers are familiar with test instruments because they have used them during their school years and at work testing products or debugging new designs. So it’s easy for them to select instruments for an automated test system. On the other hand, the engineers probably did not use switching in day-to-day testing and only consider it a minor component in an automated test system.

Choosing a switching system can be difficult because engineers often have little experience in test system design. They are accustomed to moving test probes by hand, carefully connecting the probes and ground clips to the UUT, and attaching cabling with the proper torque to ensure a quality connection. A good automated test system design should use the switching that best emulates what the engineer would have done when connecting instruments manually.

Common Problems

A common problem when implementing switching is a lack of appreciation for the potential effects of adding significant lengths of transmission lines such as cable. When making a single measurement or set of measurements with an instrument, engineers often have the test equipment and the UUT close together if for nothing more than convenience.

In an automated test system, multiple sets of measurements are made with multiple combinations of instruments. Physical-space limitations dictate that the equipment and UUT are placed farther from each other. Moreover, the addition of a switching system provides a whole other set of cabling because now the user cannot go directly from instrumentation to UUT, not to mention the length of transmission line contained within the switch system itself.

Ideally, all transmission line is lossless and has exactly the right impedance, unlimited bandwidth, and no parasitics. In practice, many limitations prevent ideal performance. Transmission lines have insertion loss, and impedance mismatch occurs at every adapter and connector interface including the switch contacts. Parasitic capacitance and inductance can limit the bandwidth and induce crosstalk between signals as well as enable coupling of noise and interference. Grounding schemes and attention to transmission-line stubs play a key role as well.

For example, Figure 1a shows a digital signal going directly from a signal generator to an oscilloscope. Figure 1b displays the same signal with a 3-ft ground wire. The waveform has significantly degraded.

As shown in Figure 1c, if the ground is forgotten or omitted for fear of ground loops, the signal loses it’s squared corners altogether. If the signal is used as part of a clocking scheme, the timing of the system may be dramatically affected if not lost.

Many times, a signal has to pass through multiple switches to get to an instrument. In those cases, special consideration must be given to transmission-line stubs of the nonclosed switches.

If the impedance and length of those stubs are not well controlled, significant degradation can occur. These stubs can become antennas or resonators. Figure 1d shows the digital signal from the example after having passed through multiple switches.

Figure 1a. Signal
Figure 1b. Signal With 3-ft Ground Wire
Figure 1c. Signal With No Ground Wire

Figure 1d. Signal After Passing Through Multiple Switches

Figure 1. Conditions Causing Digital Signal Degradation

Another very common problem encountered when selecting or designing switching systems can happen when an engineer does not ensure the switch has adequate bandwidth. Often, engineers get high-end signal generators and spectrum analyzers or oscilloscopes with wide bandwidths and then don’t give ample attention to the switching system. In those cases, the extra money for high-performance instrumentation is wasted as the switch band limits the signal before it ever gets to the UUT or measurement equipment.

For example, a well-known pacemaker manufacturer paid nearly $100,000 for its test system. With human life in the balance, no expense was spared. Then the switching system was purchased. As is often the case, little attention was paid to ensure the switch had adequate performance and would not affect the test results. Since an average heart beats around 70 times per minute and a pacemaker with defibrillator only puts out a moderate amplitude voltage signal, the customer engineer selected a low-cost switching system with 5-MHz bandwidth. The bandwidth was thought to be well beyond the 70 beats per minute (bpm).

The test engineer soon found that he was receiving an unexpected response from the device, or the device wasn’t receiving the signal at all. This was a very different result than when he connected the instrumentation and UUT directly. The engineer had forgotten or did not realize that not only is the fundamental frequency of signal important to bandwidth, but the shape of the signal is important as well.

A narrow time pulse comprises a wide range of frequencies. As a good rule of thumb, the bandwidth (BW) of a signal is approximately

BW = 0.35/tr

where: tr = rise time of the signal

The rise time of the defibrillator signal was approximately 20 ns, much faster than the 70 bpm of the fundamental signal. A 20-ns rise-time signal requires a bandwidth of 17 MHz to keep its shape.

As it turns out, the switching system that he thought well exceeded his requirements was, in fact, inadequate. Not only was the bandwidth too low, 5-MHz vs. 17-MHz requirement, but it also suffered from high levels of noise and crosstalk. All three factors were likely to have been significantly affected by insufficient attention to transmission line lengths and stubs.

Sometimes automated test systems have a very large number of signals that need to be switched to and from the UUT and instrumentation. However, there may be a limit on the number of signals that can be switched due to space limitations, interconnection difficulties, and mass cabling restraints.

With today’s advances in microelectronics, switch vendors are offering more and more dense switching modules. However, interconnecting modules to form larger switching systems still is quite a challenge. Interconnecting transmission lines that are not designed for optimal line lengths to minimize stub effects or carefully controlled to optimize match as part of the initial switching system will be problematic.

Even with more dense switching modules available and carefully selected interconnections, sometimes there may be physical restrictions on the space available for switching. In those cases, it would be advantageous to have a switching system with the flexibility to dynamically change the configuration of inputs and outputs and port-to-port connections.

Best Practices

After decades of experience in designing switching systems, engineers have developed some best practices to avoid or eliminate the common problems. These include the separation of grounds and ground planes and an uncompromising focus on signal shielding and signal isolation.

Figure 2. Switch Module

For example, Figure 2 shows what appears to be a typical switching module. However, careful attention has been given to separate the grounds for the chassis, analog circuitry, and digital circuitry, as highlighted. In addition, control, power, and signal lines have been isolated from each other by placing them on separate PCB layers interwoven between ground planes. A third step is carrying the shield between instrumentation and UUT through the switch.

The end result is that the noise and crosstalk within a switching system can be minimized, making it more likely to be transparent to the tests and measurements being made. Lower crosstalk and noise also have a direct impact on the bandwidth of the switching system.

While few switching-system designers pay such close attention to grounding schemes, shielding, and isolation, even fewer consider the basics of transmission-line theory. For example, in higher frequency switching systems, care often is given to ensure the impedance of the signal path through the switch is 50 ?. In lower frequency switching, attention to matching signal path impedance diminishes.

In both cases, the effects of the stubs created from paths not part of the signal path often are overlooked. Matching impedances ensures maximum signal integrity by minimizing reflections. Elimination, minimization, or at least careful control of stubs is critical to maximum performance of a switching system.

This approach to the signal path extends beyond the switch card. Some switching-system providers integrate a backplane into the system. The backplane allows carefully controlled transmission-line interconnections between switch modules as opposed to the possibility of a less experienced engineer interconnecting in a manner that would significantly degrade the switching system performance.

The result of implementing the best practices associated with separation of ground planes, careful attention to isolation and shielding, and a focus on transmission-line theory enables designers to develop switching systems with bandwidths typically five to 10 times greater than the majority of commercial switching systems available today. Also, these best practices provide the capability to switch lower-level signals in the presence of stronger signals and offer much better noise immunity.

The Often Overlooked Star Switch

The traditional switch topologies include the building blocks SPST, SPDT, DPDT, and multithrow switches (SPnT) and the grouping of switches to form a tree, multiplexer, or matrix. However, there is one topology that is not well known or often overlooked that can provide a switching system with many advantages: the star switch.

Figure 3a. SP4T Switch

With a multithrow switch, signals can only be routed between the common and one of the throws, positions A to D (Figure 3a). A star switch is similar to a multithrow switch without the pole or common, as shown in Figure 3b. In a star switch, the signal can be routed from any position to any other position. The SP4T has only four possible signal paths. So if a switching system was limited to using four pins, the star switch would have six possible signal paths whereas the SP4T would have only four.

Figure 3b. Star Switch

Beyond additional flexibility, the star switch also has a performance advantage. Without a common, the switch is more balanced. All of the transmission-line lengths are equal, and the stubs are shorter, allowing better impedance matching and less opportunity for noise coupling through them.

To give an example of how well balanced a star switch is, a signal with a 2-ns rise time is connected to the common node of the four-pole switch and the output of one position. The signal has a delay of about 1 ns per 8 inches.

The propagation delay between the center of the star switch and position #1 is 226 ps. The propagation delay between the center of the star switch and position #2 is 208 ps. The resultant phase difference is 18 ps. When implemented properly, the star switch can provide the ultimate flexibility and performance in a switching system even when the pin-out is already fixed.

Many military/aerospace switching systems were designed decades ago. They commonly use a form of mass interconnection to enable system resources to interact with the UUT. These are commonly referred to as a receiver mechanism. A common connection type is a pin coupled with a mating connector, so the assignment of connection on the receiver mechanism often is called the pin-out. To ensure the interface with the UUT is consistent, the pin-out usually is fixed with all connections assigned for a particular purpose and with a particular performance.

The testing and control of a UUT today require much higher bandwidths than they did decades ago. For example, many military/aerospace switching systems had less than 100 MHz of bandwidth, yet newer technologies like low-voltage differential signaling require bandwidths in excess of 500 MHz.

It is a very daunting and expensive task to swap out all of the switching systems and the interface to the UUTs. By utilizing a star switch in the switching system, not only is there much more flexibility in the interaction of system resources and UUT, but also higher performance can be attained. The pin-out can go from being fixed to being able to be dynamically reconfigured to meet both present and future requirements. In addition, the higher performance of the star switch enables the utilization of legacy-system resources while providing adaptability to future higher bandwidth or higher speed requirements.

Figure 4 illustrates the flexibility and efficiency of using a star switch architecture. In the initial design, the receiver mechanism shown in grey had a fixed pin-out, dedicating pins to specific 50-? resources, DSOs and a DMM. The switching system included nearly 30 multithrow switches.

Figure 4. Example of a Star Switching SystemPlus 10 feedthroughs to bottom resource connector(click to zoom)

By utilizing just 12 star switches, the pin-out can be reconfigured to enable the receiver mechanism to interface with any number of 50-? resources, connect to a digitizer or DMM, or route high-speed digital signals. In addition, the star switch enables the user to disconnect unwanted paths.

A very complex switch solution is achieved by using simple star switches as its core. Figure 5 shows the entire test system with the UUT and test instrumentation interconnecting through the receiver mechanism and switching system.

Figure 5. Switch Solution Using Star Switches

Conclusion

The importance of switching in an automated test system often is underappreciated. Since the switching system sits between a UUT and the test instrumentation, it dictates the performance of the test system despite well-designed products and expensive, high-end test equipment.

Common problems with ground loops and bandwidth limitation can be minimized or avoided by taking advantage of the best practices derived through decades of switching-system design experience. Separating grounds and ground planes as well as using sound transmission-line theory can produce switching systems with far greater bandwidth and isolation and far less crosstalk.

Taking it one step further, utilization of the star switch will yield ultimate flexibility in your switching system, even when your pin-out is already fixed. If you don’t have the time or expertise to take advantage of these switching system tips, contact an experienced switching solution provider. Why struggle?

About the Author

Walt Strickler is a vice president of business development for switching solutions at Giga-tronics. Mr. Strickler has more than 20 years experience in the space and defense and test and measurement industries, starting in engineering with NASA Glenn Research Center, followed by positions as worldwide sales development manager and director of marketing at Keithley Instruments, and most recently director of marketing for Bird Technologies Group. He holds a B.S.E.E. from University of Akron and an M.B.A. from Case Western Reserve University. Giga-tronics, Switching Division, 4650 Norris Canyon Rd., San Ramon, CA 94583, 925-328-4650, e-mail: [email protected]